mySpot Beamline
mySpot Beamline
The mySpot Beamline is used to provide a stable beam especially tuned for the mySpot experiment. Depending on the experiment requirements, different optical devices are used. Since the goal of the experiment is to provide several methods at the same time, beamline properties can be tuned to provide the optimal beam for a given combination of experiments. Total intensity, divergence, energy resolution, high harmonics suppression, and stability in scans can be tuned to match the requirements.
Station data | |
---|---|
Temperature range | 180 - 320 K |
Pressure range | For details contact the station manager. |
More details | mySpot |
Beamline data | |
Segment | L02 |
Location (Pillar) | 3.2 |
Source | 7T-WLS-1 (Wavelength Shifter) |
Monochromator | Multiple Crystal ( Si 111, Si 311) and Multilayer ( B4C/Mo, 2nm period ) |
Energy range | 4 - 25 keV (not all usable for all methods) |
Energy resolution | 1/8000 - 1/500 depending on used monochromator |
Flux | 1e12 - 1e13 by optimal conditions and high flux monochromator |
Polarisation | horizontal |
Divergence horizontal | 1 mrad |
Divergence vertical | 1 mrad |
Focus size (hor. x vert.) | 400x400, 400x50, further focusing in experimental hutch using capillary optics |
User endstation | not possible |
Distance Focus/last valve | Variable mm |
Height Focus/floor level | 1500 mm |
Beam availability | 24h/d |
Phone | +49 30 8062 14793 |
mySpot Beamline is used to provide stable beam especially tuned for the mySpot experiment. Depending on the experiment requirements, different optical devices are used. The schematic view shows two different configurations, one tuned for low divergency, and one for narrow energy band width, as required for the scattering and spectroscopy experiments respectively. Since the goal of the experiment is to provide several methods at the same time, beamline properties can be tuned to provide the optimal beam for a given combination of experiments. Total intensity, divergence, energy resolution, high harmonics suppression, and stability in scans can be tuned to match the requirements. For details, please visit the experiment page: mySpot.
Except 3 different monochromators, the beamline features the second, bimorph adaptive mirror, which is used to vertically focus the beam below 50µm. The mirror has 3 tracks, which are used for the suppression of higher harmonics.